WARNING: Handle lithium batteries carefully, observe
warnings on battery casing. Dispose of in accordance with local
regulations.
RECONDITION PROCEDURE
Exposure of the internal sensor to chemical vapours may interfere
with the internal sensor and cause inaccurate readings to be
logged. In a clean environment, this will slowly rectify itself.
However, exposure to extreme conditions or chemical vapours will
require the following reconditioning procedure to bring the internal
sensor back to calibration state.
80° C (176° F) at<5%RH for 36h (baking) followed by
20-30° C (70-90° F) at>74%RH for 48h (re-hydration)
High levels of pollutants may cause permanent damage to the
internal sensor.