Descargar Imprimir esta página

Data Interfaces - Sartorius YDO01M-232 Instrucciones De Instalación

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

Data Interfaces

Configuring the Data
Interface as a COM Port
(datprot)
Configure the interface as a COM port
in the Setup menu under COM1 or
UniCOM, under the "Data Protocol"
(datprot) menu item.
SBI Communication
This is a simple ASCII interface.
Data output is configured under menu
lines 6.1 and 6.3:
– Manual output of displayed value with
or without stability (menu items 6.1.1
and 6.1.2)
– Automatic output of displayed value
with or without stability (menu items
6.1.4 and 6.1.5) at intervals defined by
display updates. The number of display
updates comprising an output interval
is configured under menu item 6.3.
– Output of a configurable printout.
Output is linked to the Printout
(prtprot) settings.
If you do not activate and configure a
user-definable data record, the printout
simply contains the current value dis-
played on the indicator (weight with
unit, calculated value, alphanumeric
display).
SMA Communication
Standardized communications protocol
of the Scale Manufacturers Association
Data Input Format
(Commands)
You can connect a computer to your
display and control unit to send com-
mands controlling weighing instrument
functions and applications via the inter-
face port.
All commands use the same format
(data input format) starting with the
ESC character (ASCII 27) and ending
with a carriage return (CR; ASCII 13)
and a line feed (LF; ASCII 10). The total
length of a command is anywhere from
4 characters (1 command character
between the start and end described
above) to 7 characters (4 command
characters).
The table below shows the available
command characters; each command
must be flanked by the start and end
characters as described above.
Example: The command character for
output is "P" ("transmit readout
value"). The string "ESC P CR LF"
triggers this command.
Command Meaning
K
Weighing mode 1
L
Weighing mode 2
M
Weighing mode 3
N
Weighing mode 4
O
Block keys
P
Output readout
to data interface
R
Unblock keys
T
Tare and zero
(combination tare function)
f3_
Zero (see also the
"kZE_" command)
f4_
Tare (without zeroing;
as the "kT_" command)
F1: Trigger k key function
kF1_
F2: Trigger c key function
kF2_
(Midrics 2 only)
F3: Trigger r key function
kF3_
(Midrics 2 only)
F4: Trigger O key function
kF4_
(Midrics 2 only)
F5: Trigger w key function
kF5_
(Midrics 2 only)
Command Meaning
F6: Trigger I key function
kF6_
(Midrics 2 only)
d key
KF7_
CF: Trigger c key function
kCF_
(Midrics 2 only)
Trigger p key function
kP_
Output to printer port
Trigger T key function
kT_
(tare)
Trigger ( key function
kZE_
(zero)
x1_
Output model designation
of active weighing instrument.
Example: "LP6200S-0C"
x2_
Output serial number
of active weighing instrument;
Example: "0012345678"
x3_
Output software version
of active weighing instrument;
Example: "00-20-04"
z1_
Activate input for printout
header 1
z2_
Activate input for printout
header 2
txx...x_ xx...x: Enter letters
Length acc. to input
(Midrics 2 only)
The ASCII code for the "underline"
character is 95.
Format for entering printout header
lines: "ESC z x a ... a _ CR LF" where
x=(header line) 1 or 2, and a ... a: = up
to 20 characters of text, followed by the
"underline" character, carriage return
and line feed.
17

Publicidad

Capítulos

loading