Descargar Imprimir esta página

Installation In The Display And Control Unit - Sartorius YDO01M-232 Instrucciones De Instalación

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

4

Installation in the Display and Control Unit

Installation
Installation of the interface module in the Midrics display and control unit (and addition-
ally the installation of the cable gland and connection of the connecting cable to the ter-
minal screw strip) is only required if the Midrics display and control unit was not
equipped at the factory with this data output in accordance with the customer's order.
Notes:
§ The interface module should be installed by a certified technician who has received
specialized training from Sartorius.
§ IP65 protection:
Make sure to use the connecting cable with screw-lock hardware designed for the inter-
face module in question (see "Accessories").
!
Make sure to disconnect the equipment from power before beginning installation.
!
Any installation work that does not conform to the instructions in this manual will result
in forfeiture of all claims under the manufacturer's warranty.
!
Installation work that affects the IP65 protection rating must be performed with extreme
care.
The cable gland (IP65 protection) for connection of the interface to the display and con-
trol unit is covered by a protective cap. Please use extreme caution when performing any
work on the equipment that affects this cable gland.
§ To open the display and control unit, remove the screws from the front panel.
Installing the COM1 PCB:
§ Remove fastening screw.

Publicidad

Capítulos

loading