®
R&S
ZN-Z15
5
Operation
The external RFFE interface R&S ZN-Z15 is completely remote-controlled; it does
not provide any means for manual control.
The green LED and the orange LED on the rear panel indicate the availaility of
the regular 12V. If only the orange LED is lit this indicates insufficient input volt-
age.
The R&S ZN-Z15 is equipped with a 102x64 monochrome OLED-Display that dis-
plays the following information:
● Serial number of the instrument (for identification purposes)
● IP address
● USB and Direct Control link states
● Operational state of the instrument. This can be either
– OK if the instrument is error-free or
– ERROR <error code> with the error code of the last occurred error.
Table 5-1: R&S ZN-Z15 Error Codes
Error Code
0
3
6
9
10
11
20
22
23
24
25
User Manual 1177.5479.02 ─ 02
Description
no error
can't read from FPGA registers
config.ini corrupted (inconsistent config.ini file)
FPGA loading timeout, e.g. old FPGA not trig-
gering PROGBUS_FPGA_LOADED_GPIO
FPGA version is not compatible with µC ver-
sion
rffe.ini corrupted (inconsistent rffe.ini file)
temperature alarm FPGA
Voltage difference between -7V and +3V3 is
out of range
Voltage difference between +1V2 and Ground
is out of range
Voltage difference between +3V3 and Ground
is out of range
Voltage difference between +2V5 and Ground
is out of range
Operation
23