Environmental Parameters; Compliance With Norms - Metrix PX 110 Manual De Instrucciones

Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

4.3

Environmental parameters

I I I I I
Climatic conditions
95
90
75
45
10
-40
Note: The operating range (2) is reduced to 40°C with the mains adapter accessory.
I I I I I
Altitude
- Operation: ≤ 2,000 m
- Storage ≤ 10,000 m
I I I I I
Impermeability: protection index IP 54 (according to EN 60529, ed. 2000)
4.4

Compliance with norms

I I I I I
Electrical safety
(as per EN 61010-1 ed. 95)
- Dual insulation:
- Installation category: III
- Pollution level: 2
- Rated voltage: 600 V
I I I I I
Electromagnetic compatibility (EN 61326-1 éd.98)
I I I I I
Electrostatic discharge (IEC 61000-4-2)
- 4 kV contact, 8 kV air
- aptitude criterion A, industrial site
I I I I I
Electromagnetic field (IEC 61000-4-3)
- 10 V/m
- aptitude criterion A, industrial site
I I I I I
Bursts (IEC 61000-4-4)
- 2 kV Level III
- aptitude criterion B, industrial site
I I I I I
Shock wave (IEC 61000-4-5)
- 1 kV differential mode, 2 kV common mode
- aptitude criterion A, industrial site
I I I I I
Conducted interference (IEC 61000-4-6)
-3 V
- aptitude criterion A, industrial site
I I I I I
Mechanical resistance
- Operating position: any
- Free fall: 1 m (test as per IEC 68-2-32)
I I I I I
Auto-extinguishability
- Housing: V0
- Display window: V0
Wattmeters
0
35
50
20 26
Temperature in °C
(IP2X electrical protection for the terminals)
- Impacts: 100 g (test as per IEC 68-2-27)
- Vibrations: 0.15 mm (test as per IEC 68-2-6)
Reference range
Operating range
Storage range (without battery)
70
Chapter IV
33

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

Px 120

Tabla de contenido