CAUTION
"When checking capacitance in circuit, make sure
to turn off power and discharge capacitors fully before
making measurement.
Temperature test
Place function measuring switch to the TEMP position.
The display will show the current ambient temperatura.
Using the included type k thermocouple, connect it
using the INPUT end in the red jack and the COM end
into the black jack. Touch the object under test with
the tip of the thermocouple and the measured
temperatura will be displayed.
Frequency and duty cycle test
lnsert red probe to "INPUT' jack, and insert black
probe to "COM" jack. Place function measuring
switch to the "Hz%" position or press the "Hz%" key
in AC voltage mode. Connect the probes across the
load for measurement. Read the frequency or duty
cycle measurement on the display. Press "Hz%" to
switch between frequency and duty cycle.
Non-Contact Voltage Detection
Press NCV key. Place non-contact sensor close to the
conductor. When test voltage is greater than 90V AC
(RMS) and when the meter is close to the conductor,
the meter induction voltage indicator will flash and
buzzer will sound.
Note:
1: Even there is no indication, voltage may exist. Don't
use non-contact voltage detector to judge whether
there is voltage in the wire. Detection operation
could be affected by socket design, insulation
thickness, type and other factors.
09
2: When inputting voltage on the meter input terminal,
due to the existen ce of the induced voltage, voltage
induction indicator also may light.
3: lnterference sources of externa! environment (such
as flashlight, motor, etc.) may trigger non-contact
voltage detection by mistake.
General Specifications
General characteristics: the maximum voltage of
voltage input end and ground CAT 111 600V :":": and
600V-
Display : LCD display with maximum of 4000 counts
Measuring principie: double integral A/D convertion
Measuring range choice: Automatic
Measurement speed: 3 times/sec
Unit display: has function and power unit symbol display
Polarity indication: display"-" symbol for negativa
Overrange display: "OL"
Data hold function: display
Low battery display: display "ID" on the top of LCD
Power battery: DC1 .5V X3 SIZE AAA.
Outside measurement: 220mmx81 mmx41 mm
Weight: about 286g (include battery)
The largest size of open clamp jaw: 26mm
Operating environment: 5
Storage temperatura: -1 O
------------
polarity input
"C1"
on the top of LCD
º
º
C-35
C
º
º
C-50
C
10