CALIBRATION
This scale has been factory calibrated, and does not require calibration prior to use.
If required, the scale can be calibrated following the procedure below. Calibration of your scale
is performed using calibrated kilogram (kg) or pound (lb) weight standards. Minimum calibration
is 110 lb / 50 kg. Maximum calibration weight is 600 lb / 270 kg. Note: Only weights certified
and traceable to national standards should be used for calibration procedures.
The calibration procedure is available in the advanced settings menu. Follow this procedure to
enter the advanced menu and calibrate the scale.
1. Press
to power on the scale. Wait for "0.0" to appear on the
ZERO
display.
2. Use a small screwdriver or thin tool
to press the advanced menu button
located underneath the display
head. You may need to break
through the "SEAL" tape located
over the button.
3. When the button is pressed the display will show "SEtUP followed by
"UnIt".
HOLD
until "CAL ib" appears on the display.
4. Press
PRESS 3 SECS
FOR BMI MODE
and "CAL-U" will appear on the display.
5. Press
ZERO
6. The default calibration weight unit is kilograms. To calibrate the scale
in pounds, press
change from kg to lb. When "lb" appears on the display, press
to confirm.
HOLD
7. Press
and "CAL-0" will appear on the display.
PRESS 3 SECS
FOR BMI MODE
8. Press
. The display will show 4 or 5 digits as the Zero count
ZERO
value. Then press
display.
9. Press
HOLD
and "CAL-5" will appear on the display.
PRESS 3 SECS
FOR BMI MODE
10. Press
then enter the calibration weight. Press
ZERO
decrease the value or press
to move to the next digit..
P/N UM596KL_596KL-C REV20190923_web
Operation
to enter the weight unit. Press
ZERO
again and "CAL-0" will appear on the
ZERO
TARE
to increase the value. Press
BSA
HOLD
PRESS 3 SECS
to
FOR BMI MODE
HOLD
to
PRESS 3 SECS
FOR BMI MODE
17
Display
"0.0"
"SEtUP"
followed by
"UnIt"
"CAL ib"
"CAL-U"
"kg" or "lb"
ZERO
"CAL-0"
"XXXXX"
followed by "CAL-
0"
"CAL-5"
"XXX"