Container systems

5.6 Filter change

For removing the filter holder, the safety catch must be pushed upwards and then the holder
must be pulled in the direction of the arrow (see Fig. 2).
After placing the filters over the perforated areas on the inside of the lid and (if applicable) on
the bottom of the container, the lock have to be pressed down until the filter holder snap into
position. HEBUmedical lids should only be assembled with HEBUmedical filter holders.
The disposable paper filters have to be exchanged before every new sterilisation
cycle.
Only HEBUmedical filters can ensure suitability and exactness of fit.
Guarantee can only be accepted when original HEBUmedical filters have been used
!
Only original HEBUmedical component parts such as lids, bottoms,
filters, seals and filter holders should be combined with each other
so that the container sealing and function as a germ barrier is not
compromised. All filters or seals made by other manufacturers must
be approved by HEBUmedical before used with the HEBUmedical
container-system. Otherwise warranty exclusion will be the result.

6 Loading

The overall weight for loading containers should not exceed the following recommended
loads, as otherwise satisfactory sterilisation of sterile materials cannot be ensured.
Model
Dimension
Fig. 2: Filter holder with a paper filter
Maximum Recommended Load in kg
Instruments
EN
Textiles
25