Descargar Imprimir esta página

HEBU medical HE 3110-11 Instrucciones Para El Uso página 26

Publicidad

Idiomas disponibles

Idiomas disponibles

EN
1/1 (Full-) Size
Container
¾ Size-
Container
½ Size-
Container
Mini-
Container
Dental-
Container
When loading with textiles, pay attention that the folded textile or laundry articles are
positioned vertically. In the case of a full container, it should be possible to push an
outstretched hand between the pieces of laundry without effort
The sterilisation of various container loads and packaging
configurations has to be validated by the responsible specialist
!
Complex instruments like Endoscopes, instruments with a lumen,
compressed air-driven instruments or power systems and
instruments with cannulae are to be prepared according to the
manufacturers' instructions for sterilisation and to be sterilised
accordingly. Small baskets, trays, other types of accessories,
especially with cover or lids, should only be used with the
sterilisation container, if the sterilisation container has been
Please see details about HEBUmedical frames, baskets, and metal
!
When using non-absorbent tray liners (e.g., plastic/silicone-fingered
organising mats) for the containers this can cause condensate to
pool. Instead of this use moisture absorbing mats if necessary.
26
580x280x110
580x280x160
580x280x210
580x280x260
465x280x110
465x280x160
465x280x210
285x280x110
285x280x160
285x280x210
285x280x260
300x140x40
300x140x70
300x140x100
310x190x40
310x190x65
310x190x130
hygiene personnel according to ISO 11134.
specifically designed and tested for that purpose.
sterilising trays in our product catalogue.
3,8
5,8
7,7
10
3,1
4,6
5,9
1,9
2,8
3,8
4,9
0,4
0,7
1
0,6
0,9
1,8
Container systems
3
4,6
6,1
8
2,5
3,7
4,7
1,5
2,2
3
3,9
0,3
0,5
0,8
0,5
0,7
1,4

Publicidad

loading