DisinfectDaily:
1.
U singacleancontainerorbowl,soakitemsinthree
partswarmwatertoonepartwhitevinegarfor30
minutes(Fig.8)ORuseamedicalbacterial-germicidal
disinfectantavailablethroughyourprovider.Besureto
followmanufacturer'sinstructionscarefully.
2.
W ithcleanhands,removeitemsfromdisinfectant
solution,rinseunderhottapwater,andairdryona
cleanpapertowel.Storeinazip-lockbag.
NOTE– Do not towel dry nebulizer parts; this could cause
contamination.
Topreventpossibleriskofinfectionfromcontaminatedcleaningsolutions,
alwayspreparefreshsolutionforeachcleaningcycleanddiscardsolution
aftereachuse .
3.
K eeptheoutersurfaceofthetubingdust-freebywipingregularly.Nebulizertubing
doesnothavetobewashedbecauseonlyfilteredairpassesthroughit.
NOTE– The disposable nebulizer will last at least 15 days and possibly longer, depending on
usage. Proper cleaning will also help extend the life of the nebulizer. Because it is
disposable, we recommend that an extra nebulizer be kept on hand at all times. The
reusable nebulizer is dishwasher safe and may be cleaned and reused for up to one
year.
ComPRESSoR ClEANING
1.
W ithpowerswitchinthe"Off"position,unplugpowercordfromwalloutlet.
2.
W ipeoutsideofthecompressorcabinetwithacleandampclotheveryfewdaysto
keepdust-free.
D onotsubmergeinwater;doingsowillresultincompressordamage .
FIlTER CHANGE
1.
F iltershouldbechangedevery6monthsorsooneriffilterdiscolors.
2.
R emovefilter(Fig.A)bygraspingitfirmlyandpullingoutoftheunit.
Discardoldfilter.
3.
R eplacewithanewDeVilbissfilter(part#3655D-601).Additionalfiltersshouldbe
purchasedfromyourDeVilbissprovider.
4.
P ushnewfilterintoposition.
CAUTION– Do not reuse or wash filter or substitute any other material such as
cotton for a DeVilbiss air-inlet filter. It will result in compressor damage.