Recommended Cleaning Intervals; Product Disposal; Troubleshooting - DeVilbiss Healthcare SleepCube STANDARD CPAP DV51 Serie Guía De Instrucciones

Dispositivo de presión positiva en las vías respiratorias
Ocultar thumbs Ver también para SleepCube STANDARD CPAP DV51 Serie:
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

Recommended Cleaning Intervals:

CPAP outer housing
Tubing
Gross Particle Intake Filter 10 days
Fine Particle Intake Filter
Humidifier cradle
Humidifier chamber
Humidifier chamber
5 years @ 7 day interval = 261 cleaning cycles
* Number of cleaning cycles determined by recommended cleaning interval and expected service life
** Only weekly dishwasher cleaning due to the high temperature of the water and the high concentration of the soap.
Mask and Headgear
Clean the mask and headgear according to the manufacturer's instructions.

PRODUCT DISPOSAL

The outer packaging is made of environmentally friendly materials that can be used as secondary raw materials. If you no
longer need this packaging, bring it to your local recycling and waste disposal facility according to the applicable
regulations for your location.
The equipment, including accessories and internal components, does not belong in your regular household waste. Such
equipment is manufactured from high-grade materials and can be recycled and reused. The internal motor, PC board,
power supply board and wiring should be removed from the unit and recycled as electronic waste. The remaining plastic
components should be recycled as plastics.
The European Directive 2012/19/EU on Waste Electrical and Electronic Equipment (WEEE) requires that electrical and
electronic equipment be collected and disposed of separately from other unsorted municipal waste, with the aim of
recycling it. The crossed out waste bin symbol indicates that separate collection is required.

TROUBLESHOOTING

WARNING
Electric Shock Hazard - Do not remove the device cabinet. There are no user-serviceable internal components.
The cabinet should only be removed by a qualified DeVilbiss provider.
Issue
Nothing shows on
the display.
EN - 12
Recommended
cleaning interval
7 days
Daily
Do not clean, replace if dirty N/A
1 / week
1 / day
1 / week
Possible Cause
1. Device is not plugged in, or the power cord
is not fully inserted.
2. Optional Humidifier-The flow generator is
not fully seated on humidifier cradle.
3. There is no outlet power.
Number of cleaning
cycles *
261
180
18
261
365
52
Remedy
1a. Verify that the power cord is firmly connected
to the SleepCube and the power outlet.
1b. If you are using a DC power source, make
sure the cable connections are secure. Verify that
battery is charged.
2. Refer to assembly instructions (Fig. 1-7).
3. Locate a suitable functioning power source.
Compatible cleaning agents /
max washing temperature
Water, use only a damp cloth
Mild dish soap and warm water
Mild dish soap and warm water
N/A
Water, use only a damp cloth
Mild dish soap and warm water
Dishwasher **, top shelf only
SE-DV51-1

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

Intellipap dv51 standard cpap serieDv51 standard cpap serie

Tabla de contenido