SECTION 7—MAINTENANCE
Nebulizer Change
The nebulizer is designed for single patient use.
The nebulizer MUST be replaced after a long period of inactivity or when
it is obstructed by dry medication, dust, etc. The reusable nebulizer should
last from six months to one year with general use. The disposable
nebulizer should last 30 days. For the nebulizer that is disposable, it is
recommended that an extra nebulizer be kept at all times.
Filter Change
1.
The filter MUST be replaced after approximately 30 hours of use or
when it turns grey.
2.
Open the filter holder to replace the filter with a new one.
DO NOT use cotton or any other material.
DO NOT wash or clean the filter.
Use only Invacare filters.
DO NOT operate unit without an air filter.
Stratos™Compact
CAUTION
24
Part No 1130203