Mini Neb /English
6.3 Fuse Change
NOTE –
Use identical fuse with correct rating recommended by manufacturer.
1.
Use a coin; unscrew the fuse cover in the bottom of
the device.
2.
Remove the fuse from the fuse cover.
3.
Insert new fuse with correct rating recommended by
manufacturer.
4.
Screw the fuse cover back into position.

7. Troubleshooting

If your MiniNeb Compressor Nebulizer fails to function, consult the Troubleshooting guide
below. If the problem persists, consult your equipment provider.
Problem
Cause and Solution
Device
doesn't
operate.
Weak
Nebulization
Check if plug is properly fit into an appropriate electrical outlet.
When device has been run continuously for over 30 minutes right before
using, an auto shut down may activate by built-in thermal protector, cool
down device for 30 minutes before next usage.
Check if fuse is blown and properly situated. (see "Maintenance" section
for details)
Check for proper electrical voltage.
Check tubing for blockage or air leakage at connection to MiniNeb
compressor or nebulizer cup, replace as needed.
Check the nebulizer cup if it is properly assembled and not damaged. If
there is any damage, replace as needed.
Check if filter is too dirty, replace as needed.
9