clar N and ri-clar
®
55/100 mm
ri-clar
®
2.3.6 Attaching the headlight / the head mirror
ri-focus
®
LED
The lamp head is firmly fixed to the headband and can only be exchanged together with the headband.
ri-focus
®
ri-focus
®
reverse ri-clar
clar N
Attention!
• Never touch the lamps during use.
They can get very hot!
• In the case of models ri-focus® and ri-focus® reverse, only touch the ring at the front of the lamp head or the grip at the rear
of the lamp head during operation. All other parts can heat up considerably.
• In the case of the Clar models, you can touch the following parts during operation: the plastic shell on the mirror, the adju-
sting knob and the swivel arm.
16
clar N
®
Push the lamp head into the mounting on the front of the headband.
By opening the plastic screw, the clamping unit located on the headband is opened up until the ball-
and-socket joint on the mirror fits. Attach the mirror by turning the plastic screw tightly on the head-
band.
Attention!
Before exchanging the complete mirror, the electric plug located on the joint of the lamp has to be
pulled out. After the mirror has been exchanged, the plug can be pushed back in again.
If the lamp does not function after plugging the cable back in again, the plug has to be rotated 180°
and plugged in again.
Manually moving the lamp fixture towards and away from the
mirror.