Blood Glucose Monitoring Systems – User's Manual
P R O D I G Y
®
Step 2
– Apply Control Solution
Shake the control solution bottle well.
Remove the cap. Squeeze the bottle, discard
the first drop, and wipe off the dispenser tip
with a clean tissue paper or cotton. Squeeze
the bottle again to get a second drop onto a
clean non-absorbent surface or on your fin-
gertip first. Then apply the drop to the test
strip absorbent channel, (where it meets the
narrow channel) until the drop is drawn into
the test strip and the confirmation window
Figure 15
is filled. The meter begins to count down.
Caution: To avoid contaminating the control solution with the
content of the test strip, you have to place a drop of control solution
on a clean surface or on your fingertip first. Then touch the strip to
the drop. DO NOT DIRECTLY APPLY CONTROL SOLUTION ONTO A STRIP.
Incorrect – Figure 17
– Figure 16
Correct
19
Diagnostic Devices, Inc.
www.prodigymeter.com
1- 800-243-2636