Step-by-Step Test Procedure

Step 1
– Insert the Test Strip
Insert a test strip with contact bars end first into the
test slot. (Contact bars must be inserted all the way
into the meter or you may get an inaccurate test
result.) The meter turns on automatically.
Step 2
– Apply Sample
For PRODIGY Audio and PRODIGY
Autocode Talking only. You will hear a voice message,
"Please apply blood into the strip".
When the meter shows " ", obtain a drop of blood
using the adjustable Lancing Device. The blood
sample must be at least 0.6 microliter in volume
(
actual size) or you may get an inaccurate test
result. Apply blood to the strip absorbent channel
where it meets the narrow channel. Do not push your
finger against the test strip or try to apply a smeared
sample. The Test Strip confirmation window should be
completely filled before the meter begins to count
down. If not, do not try to add more blood to the test
strip. Discard the test strip and retest with a new one.
If you have trouble filling the test strip, please contact
Technical Support for assistance.
Correct – Figure 27a
24
Diagnostic Devices, Inc.
Blood Glucose Monitoring Systems – User's Manual
®
www.prodigymeter.com
Incorrect – Figure 27 b
Figure 26
Figure 28
1- 800-243-2636