Insert tubes into ColorQ with line and arrow facing forward.
Eng
!
"Sun" = Shade meter from bright sunlight. Press/Hold button to turn off.
! "Er7" = Tube is misaligned, scratched, stained, or light chamber is stained.
See User Guide for more instructions.
! Brush and rinse all tubes promptly after use.
! Allow tubes to dry before returning to case.
! Replace reagent caps securely.
!
If unit should auto-off during testing rinse and fill one tube to Blank and return
to testing.
!
Use only LaMotte replacement reagents.
!
bAt = Battery is low.
!
Insertar el tubo en el ColorQ recto.
Es
s
!
"
" = apartar el aparato de los rayos del sol. Presionar/mantener el botón
para apagar.
! "Er7" = El tubo está mal colocado o rayado, o la cámara de lectura está
manchada. Para más instrucciones, consulte la Guía del usuario.
!
Si el aparato se auto-apaga durante el test, llenar un tubo para el Blanco y
volver al análisis.
!
Enjuagar y limpiar todos los tubos adecuadamente después del uso.
!
Permitir que los tubos se sequen antes de volver a usarlos.
!
Cerrar correctamente los reactivos.
!
Utilice unicamente los reactivos LaMotte. (tabletas grado instrumento).
!
bAt = Batería baja.
!
Introduire le ColorQ droit.
s
Fr
!
"
" = Maintenir l'appareil à l'ombre.
Appuyer le bouton off pour l'eteindre.
! "Er7" = Le tube est mal placé ou rayé, ou
la caméra est tachée. Voir le Guide de
l'utilisateur pour toute autre instruction.
!
Si l'appareil s'éteint automatiquement refaire
un blanc et reprendre l'analyse.
!
Essuyer et nettoyer tous les tubes
correctement après l'usage.
!
Permettre que les tubes se sechent bien
avant de les réutiliser.
!
Ferrer correctement les reactifs.
!
Utilisez des réactifs LaMotte pour rechanger
uniquement.
!
bAt = Batterie basse.
Q
Pro-11
For software version 2.00 or higher
PO Box 329 • Chestertown • Maryland • 21620 • USA
800-344-3100 • 410-778-3100 • 410-778-6394
www.lamotte.com
Order Code 2058
WARNING! This set contains chemicals
that may be harmful if misused. Read
cautions on individual containers carefully .
Not to be used by children except under
adult supervision.
03/19
62058-V2

Publicidad

loading

Resumen de contenidos para LaMotte ColorQ Pro-11

  • Página 1 Enjuagar y limpiar todos los tubos adecuadamente después del uso. Permitir que los tubos se sequen antes de volver a usarlos. Cerrar correctamente los reactivos. Utilice unicamente los reactivos LaMotte. (tabletas grado instrumento). bAt = Batería baja. Introduire le ColorQ droit.
  • Página 2 “Blank” dans l'appareil. Aller au chlore (FCL). Enlever le tube. * Reagent is a potential health hazard. Read SDS at www.lamotte.com. * Los reactivos son peligrosos para la salud. Leer SDS en www.lamotte.com. * Les Reactifs sont dangereux pour la santé. reactivos son peligrosos. Lire SDS en www.lamotte.com.
  • Página 3 ColorQ Range: 0-10 ppm Chlorine, FREE CHLORINE/BROMINE 0-22 ppm Bromine fcl / DPD 1A - Code P-6740-G *DPD 1B - Code P-6741-G Color 1. Add 5 drops of DPD 1A and 5 drops of DPD 1B to same tube. 2. Cap and mix. Insert tube into ColorQ. 3.
  • Página 4 TOTAL CHLORINE ColorQ Range: 0-10 ppm Chlorine, 0-22 ppm Bromine DPD 3 - Code P-6743-G Color Color 1. Remove cap from reacted FCL tube (Free Chlorine). 2. Add 5 drops of DPD 3 to the tube. 3. Cap and mix. Insert tube into ColorQ. 4.
  • Página 5 ColorQ Range: 6.5-8.5 pH - Code 7037-G Color 1. Fill a clean tube to the 5mL line with water sample. 2. Add 5 drops of 3. Cap and mix. Insert tube into ColorQ. 4. Press button to read “PH” pH. Remove tube. 1.
  • Página 6 ALKALINITY ColorQ Range: 0-250 ppm ALK - Code 7038-G Color 1. Press button to go to “ALY” Total Alkalinity. 2. Fill a clean tube to the 5mL line with water sample. 3. Add 5 drops of 4. Cap and mix. Insert tube into ColorQ. 5.
  • Página 7 CALCIUM HARDNESS ColorQ Range: 0-700 ppm *CH1 - Code 7042-G for use with ver 1.40 or greater CH2 - Code 7041-G Color 1. Press button to go to “CH” Calcium Hardness. 2. Fill a clean tube to the 5mL line with water sample. 3.
  • Página 8 CYANURIC ACID (STABILIZER) ColorQ Range: 0-125 ppm *CYA - Code 6996A-J Color 1. Press button to go to “CYA” Cyanuric Acid. 2. Fill a clean tube to the 5mL line with water sample. 3. Add 1 CYA tablet. Crush with tablet crusher. Cap. 4.
  • Página 9 COPPER ColorQ Range: 0-5 ppm * Copper - Code 3808A-H Color 1. Press button to go to “COP” Copper. 2. Fill a clean tube to the 5mL line with water sample. 3. Add 1 COP BCA tablet. Crush with tablet crusher. Cap. 4.
  • Página 10 IRON ColorQ Range: 0-3 ppm *Ir - Code 3725A-J Color 1. Press button to go to “Ir” Iron. 2. Fill a clean tube to the 5 mL line with water sample. 3. Add 1 IRON Tablet. Crush with tablet crusher. 4.
  • Página 11 BIGUANIDE ColorQ Range: 0-70 ppm Biguanide Test Solution Code 3919-G Color 1. Press button to go to “bi 9” Biguanide. 2. Fill a clean tube to the 5 mL line with water sample. 3. Add 5 drops of BIGUANIDE TEST SOLUTION. 4.
  • Página 12 BIGUANIDE SHOCK (PEROXIDE) ColorQ Range: 0-160 ppm b 5 h Biguanide Shock - Code 6930A-J Color 1. Press button to go to “b Sh” Biguanide Shock. 2. Fill a clean tube to the 5 mL line with water sample. 3. Add 1 SHOCK Tablet.