Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

SAFETY PRECAUTIONS
Always keep the sound head in full contact with the patient' s skin or
submerged under water when setting intensity.
Use ample conductive gel to ensure good coupling throughout the
treatment. If needed, apply when setting intensity.
Use of controls, adjustments, or performance of procedures other than
those specified herein may result in hazardous exposure to ultrasonic
energy.
Do not apply the Ultrasound Applicator to the patient during the Head
Warming period. Applicator must remain in Applicator Hook duing the
Head Warming period.
Do not drop the applicator on hard surfaces. Do not cool an overheated
sound head with ice water or ice packs. Do not allow the sound head
to reach maximum temperatures repeatedly. All of these conditions are
likely to damage the sound head crystal. Damage resulting from these
conditions is not covered under the warranty.
Be sure to read all instructions for operation before treating a patient.
Dispose of all products in accordance with local and national regulations
and codes.
Use of controls, adjustments, or performance of procedures other than
those specified herein may result in hazardous exposure to ultrasonic
energy.
Use only accessories that are specially designed for this unit. Do not use
accessories manufactured by other companies on this unit. DJO, LLC
is not responsible for any consequence resulting from using products
manufactured by other companies. The use of other accessories or cables
may result in increased emissions or decreased immunity of this unit.
Patients with an implanted neurostimulation device
must not be treated with or be in close proximity to
any shortwave diathermy, microwave diathermy,
therapeutic ultrasound diathermy, or laser diathermy
anywhere on their body. Energy from diathermy (shortwave,
microwave, ultrasound, and laser) can be transferred
through the implanted neurostimulation system, can cause
tissue damage, and can result in
severe injury or death. Injury, damage, or death can
occur during diathermy therapy even if the implanted
neurostimulation system is turned "off. "
DO NOT connect the unit to an electrical supply without
first verifying that the power supply is the correct voltage.
Incorrect voltage may cause unit damage, malfunction,
electrical shock, fire, or personal injury. Your unit was
constructed to operate only on the electrical voltage
specified on the Voltage Rating and Serial Number Plate.
Contact your DJO, LLC dealer if the unit is not properly
rated.
5
Intelect® Legend Series Combo

Publicidad

Tabla de contenido
loading

Productos relacionados para Intelect COMBO LEGEND Serie

Este manual también es adecuado para:

Combo 2cCombo 4c

Tabla de contenido