Figure 9 Power connections
1 Conduit (preferred)
4.3.3 Wiring for power
To connect the instrument to power, refer to
Figure 10 Power connection
1 Voltage switch (in 115 V position)
2 Voltage switch (in 230 V position)
Wire color code
North America
International
electrotechnical
commission (IEC)
4.3.4 Select the voltage
To prevent serious damage to the instrument, make sure that the line voltage is set correctly. Refer to
on page 14.
14 English
Figure 10
Table 1 Terminal wiring
Protective earth ground
Green
Green with yellow line
N O T I C E
2 Power cord (optional)
and
Table
1.
3 AC power connector
4 Fuses (F1 and F2)
Hot or Ø1
Black
Brown
Neutral or Ø2
White
Blue
Figure 10