Setting Reflected Temperature Compensation
To enable and set reflected temperature compensation, use the main program
menu to select:
[Setup] → [Reflection Compensation]
Reflected
[Compensate
temperature value can be set
between -40 and 1000°C.
OPERATION
Once the sensor is in position and the appropriate power, air, water, and cable con-
nections are secure, the system is ready for continuous operation by completing the
following simple steps:
1. Turn on the power supply
2. Turn on the instrument or PC
3. Read / monitor the temperature
IMPORTANT
Be aware of the following when using the sensor:
• If the sensor is exposed to significant changes in ambient temperature (hot to
cold, or cold to hot), allow 20 minutes for the temperature to stabilise before tak-
ing or recording measurements.
• Do not operate the sensor near large electromagnetic fields (e.g. around arc
welders or induction heaters). Electromagnetic interference can cause measure-
ment errors.
• Wire must be connected only to the appropriate terminals.
MAINTENANCE
Our customer service representatives are available for application assistance, cali-
bration, repair, and solutions to specific problems. Contact our Service Department
before returning any equipment. In many cases, problems can be solved over the
telephone. If the sensor is not performing as it should, try to match the symptom
below to the problem. If the table does not help, call I-Tec for further advice.
TROUBLESHOOTING
Symptom
No output
Erroneous temperature
Erroneous temperature
Erroneous temperature
LENS CLEANING
Keep the lens clean at all times. Any foreign matter on the lens would affect meas-
urement accuracy. Blow off loose particles (if not using the air purge accessory)
with an air 'puffer'.
GUARANTEE
I-Tec guarantees each instrument it manufactures to be free from defect in material
and workmanship under normal use and service for the period of two years from the
date of purchase. This guarantee extends only to the original buyer according to
›
I-Tec Terms and Conditions of Sale.
8
temperature
compensation
for Reflected Temperature], and when enabled the
Probable Cause
No power to sensor
Incorrect wire connection
Faulty sensor cable
Field of view obstruction
can
be
enabled/disabled
Solution
Check power supply
Check wire colour codes
Verify cable continuity
Remove obstruction
using