HEINE NT300 Manual De Instrucciones página 13

Ocultar thumbs Ver también para NT300:
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 24
Maintenance
The charger is maintenance free. However, it must be tested by a qualified electrician at periodic intervals of
24 months.
Do not open the power adapter (risk of electric shock).
Service
This product does not require regular service.
Disposal
The product must be recycled as separated electrical and electronic devices. Please observe the relevant
state-specific disposal regulations.
Used or defect batteries must be disposed adequately. Avoid skin contact with defect batteries.
Technical specification
Mains supply:
Current consumption:
Output:
Fuse:
Charging current:
Charging time:
Safety class:
Protection class:
Environmental conditions for
operation
Temperature:
Humidity:
Air pressure:
Environmental conditions for
transport and storage
Temperature:
Humidity:
Air pressure:
General Notes and Warnings
The HEINE
NT300 charger is a high-quality piece of equipment. Always handle with care.
®
Check the correct operation of the product at regular intervals. Do not use the equipment if it shows
visible signs of damage.
Do not use the device in hyperbaric chambers, in explosive or oxygen loaded environments!
Do not use the device near strong magnetic fields like MRI scanners.
We can only guarantee the performance of this device when fitted and used with original
HEINE accessories and spare parts. Otherwise the HEINE Warranty is terminated.
The NT300 must not be charged using rechargeable batteries from other manufacturers.
Do not manipulate or modify the product!
The LED indicates that the power adapter is ready for use.
Do not pull the cable to disconnect the power adapter from the mains.
Do not store or use the charger outdoors.
100 – 240 V AC / 50 – 60 Hz
max. 30 VA
15Vdc, 2A
Integral overload protection
Max. 1.5 A for BETA L
Max. 0.5 A for other HEINE rechargeable batteries
BETA L rechargeable battery -> approx. 2 hrs
Other HEINE rechargeable batteries -> up to 4 hrs
II
IP 20
+10°C to +40°C
10% to 75% r.h.
700 hPa to 1060 hPa
-20°C to +50°C
10% to 95% r.h.
500 hPa to 1060 hPa
med 0613
2014-03-10
13/76

Publicidad

Tabla de contenido
loading

Productos relacionados para HEINE NT300

Este manual también es adecuado para:

Beta tr

Tabla de contenido