Dust Ignition Proof (FM 3010145)
DIP Class II, Division 1, Groups E, F and G
DIP Class III, Division 1
Non Incendive (FM 3010145)
NI Class I, Division 2, Groups A, B, C and D
Environmental Protection (FM 3010145)
Option: Type 4X or Type 4
Special conditions for safe use:
Entity Parameters:
Vmax = 30 Vdc, Imax = 110 Ma, Ci = 5 nF, Li = 12 uH
Temperature Class: T4
Maximum Ambient Temperature: 60ºC (-20 to 60 ºC)
Explosion Proof (Nemko 01ATEX445) - In Progress
Group II, Category 2 G, Ex d, Group IIC, Temperature Class T6, EPL Gb
Maximum Ambient Temperature: 40ºC (-20 to 40 ºC)
Environmental Protection (Nemko 01ATEX445)
Options: IP66/68W or IP66/68
The transmitters are marked with options for the indication of the protection code. The certification is valid only when the
protection code is indicated in one of the boxes following the code.
The Essential Health and Safety Requirements are assured by compliance with:
EN 60079-0:2009 General Requirements
EN 60079-1:2007 Flameproof Enclosures "d"
Intrinsic Safety (DMT 00 ATEX E 085) - In Progress
Group I, Category M2, Ex ia Group I, EPL Mb
Group II, Category 2 G, Ex ia, Group IIC, Temperature Class T4/T5/T6, EPL Ga
Supply and signal circuit designed for the connection to an intrinsically safe 4-20 mA current loop:
Ui = 28 Vdc, Ii = 93 mA, Ci ≤ 5 nF Li = Neg
Maximum Permissible power:
Ambient Temperature: -40ºC ≤ Ta ≤ + 85ºC
The Essential Health and Safety Requirements are assured by compliance with:
EN 60079-0:2009 General Requirements
EN 60079-11:2007 Intrinsic Safety "i"
EN 60079-26:2007 Equipment with equipment protection level (EPL) Ga
NEMKO (Norges Elektriske MaterielKontroll)
EXAM (BBG Prüf - und Zertifizier GmbH)
Max. Ambient temperature Ta
85°C
75ºC
44ºC
50ºC
55ºC
60ºC
65ºC
70ºC
40ºC
Temperature Class
Power Pi
T4
700 mW
T4
760 mW
T5
760 mW
T5
700 mW
T5
650 mW
T5
575 mW
T5
500 mW
T5
425 mW
T6
575 mW
Informaciones Sobre Certificaciones