Idiomas disponibles

Idiomas disponibles

4.3. - TURNING ON THE LAMP
1. Turn on the switch of the room. A red light should appear on the keypad of the lamp while it is connected.
2. Turn on the lamp, using the keypad nº 5 and adjust the light intensity with the + and – buttons. (See Fig. 3)
3. To change the position of the lamp, use the handle nº 1. Do these gently avoiding brusque movements.
4. The lamp is adjusted at factory. If necessary a readjustment, see point 6.2.
Fig. 3 – Keypad
4.4.- RECOMENDATIONS FOR USE
Move the lamp gently, avoiding brusque movements.
If the lamp is not in use do not leave it on, since apart from wasting electricity you will reduce the useful life of the LED.
5.
- CLEANING
Before doing anything, make sure the lamp is off.
The lamp must be cleaned regularly with a cloth dampened with water, taking care that water does not reach the inside of
the equipment via the ventilation grille. For persistent stains such as blood, disinfectants, etc. a little neutral soap can be
used. Do not use products containing acetone in any case. Dry the lamp well before use. If there is liquid inside you
must wait until it has evaporated.
The handle located on the cupola is sterilizable by steam media at a temperature of 121°C with an exposure time of 20
minutes.

6. - MAINTENANCE AND UPKEEP

6.1. - PERIODIC CHECKS

According to hospital procedures, the maintenance service should periodically carry out the following checks:
1. Check the general working order of the lamp.
2. Check the mobility of the arm and cupola. See point 6.2.
3. Check the equipment is still earthed and is correctly connected.
4. Check its general state (of cleanliness, etc...).
6.2. - ADJUST OF THE ARM
Check the adjustment of the round joints, both of the
rocker arm and the fork and the dome. If necessary,
proceed to adjust them: it must move effortlessly and
stay in position. An excessive tightening of them may
produce a blockage of the movement. Adjust by
screws nº 18, 19, 20 and 21.
6.3. - CHANGE THE FUSES
If the switch is "on" and the lamp is connected to the mains but it does not work, then check the fuse following these
instructions:
A – On / Off button
B – On / Off Led indicator
C – Luminous intensity adjustment buttons
D – Luminous intensity Leds indicators.
Fig. 4a – Arm adjustment.
Fig. 4b – Cuppola and fork adjustment.