Hygienic Preparation; Time Periods - Weinmann OXYMAT 3 Manual De Instrucciones

Ocultar thumbs Ver también para OXYMAT 3:
Tabla de contenido

Publicidad

No water or disinfectant must be allowed
to enter the air slits or the tubes.
Vatten eller desinfektionsmedel får ej
tränga in i ventilationsöppningarna eller
slangarna.
En la rejilla de ventilación y en los tubos
flexibles no debe entra agua ni desinfec-
tante.
98

6. Hygienic preparation

The OXYMAT 3 and its accessories must al-
ways be subjected to hygienic preparation at
regular intervals. After performing hygienic
preparation please always carry out a func-
tional check (see "7. Functional Check"
page 106).

6.1 Time periods

Cleaning of the device and accessories must
be performed at the intervals given below.
We recommend also carrying out disinfection
within these periods.
After every usage:
– nasal cannula
– connecting tubes
Daily:
– sterile water tank
Weekly:
– dosage monitor
– flowmeter
– coarse dust filter
– OXYMAT 3 housing
– power cord
In the case of patients with a weakened im-
mune system or a special history of illness dis-
infection of the individual components may be
necessary after every usage. Please consult
your doctor in this case.
Englisch

Publicidad

Tabla de contenido
loading

Tabla de contenido