Electrical Installation - Williamson SilverC Serie Manual De Instalación Y Funcionamiento

Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

air/Water cooleD housing
The air/water cooled housing shown below allows the sensor to withstand high ambient temperatures.
It is equipped with two 1/8" NPT fittings with 1/8" BSP fittings availble as an option. Water temperature
should be 50°F (10°C) to 80°F (27°C) for efficient cooling. Chilled water below 50°F (10°C) is not recom-
mended. To avoid condensation, the air purge collar should be used with the water-cooled housing.
1/8 NPT water/air connections
0.75
2.5 (63)
(19)
4.1 (103)
Legend = inches (millimeters)
air purge collar
The air purge collar below is used to keep dust, fumes, moisture, and other contaminants away
from the lens. It must be screwed in fully. Air flows into the 1/8" NPT fitting (1/8" BSP fittings
availble as an option) and out of the front aperture. Air flow should be no more than 1 to 3 cfm (0.5
to 1.5 litres/sec).
Clean or 'instrument' air is recommended.
1/8 NPT Air connection
1.0 (25)
AP1
2.0 (50)
AP2
Legend = inches (millimeters)
6
M16 x 1mm
0.83
(21)
1.57 (40.0)
0.79 (20)
1.14 (29)
1.57 (40)

electrIcal InstallatIon

SilverC9 Sensor
Two-wire units (option 0)
SilverC9 Sensor
Four-wire units (options 1 to 4)
SilverC9A Sensor
Four-wire units (option 0)
Display/Controller
Power Supply
+
-
PWR+
-
PWR-
+
4-20mA
4-20mA = Target
Temperature
Display/Controller
Power Supply
+
-
PWR+
-
PWR-
+
4-20mA
4-20mA = Sensor
Temperature
Display/Controller
OP-
-
OP+
+
mV/TC = Target
Temperature
Display/Controller
Power Supply
+
-
PWR+
-
PWR-
+
4-20mA
4-20mA = Target
Temperature
EA Module or PLC
Power Supply
-
IP-
IP+
4-20mA
+
-
SC 0V 24V
4-20mA = Emissivity
Setting
+
7