User Replaceable Accessories And Spare Parts - VWR VisiCam Serie Manual De Funcionamiento

Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 39
Possible reasons:
1. The camera USB cable is not well connected. -- Reconnect the camera again.
2. Check whether the camera is connect to the USB port on the front of the mainframe. If so, then
connect it to the USB port on the back of the mainframe. -- The USB ports on the front of the mainframe
are the extended ports from the motherboard which can not provide stable voltage and current for
cameras.
3. If the camera itself needs extra power supply—check if the power supply has been connected
and switched on.
4. The camera driver is not installed or failed to install. -- Go to the 'Device Manager'--> 'Imaging
Devices' and check whether the camera is listed WITH YELLOW FLAG. Then re-install the driver again.
If still fail to install driver, please contact camera technical support for help.
On preview images, lots of horizontal lines appear
Colours in Preview window are not properly displayed.
(1)Perform white balance with microscope lamp switched on and without any specimen under microscope
objective
The image in the Preview window appears overexposed even if the camera exposure is set at
minimum
(1)Insert grey filters on the microscope illuminator to reduce the amount of light reaching the camera. If you
reduce illumination level of microscope, consider that colour regulation could be altered and you will have
to perform again a white balance.
Maintenance
When not in use, please store the camera in a dry place and far from hot sources.
Avoid touching the sensor/filter surface with fingers, and other objects. This may damage the sensor or let
dust go over it.
To clean the filter over the sensor it is enough to use a swab stick with 100% alcohol
Do not try to open the camera by yourself, this will void the warranty

User replaceable accessories and spare parts

Description
C-mount Adapter for Visiscope TL324
C-mount Adapter for Visiscope TL524
C-mount Adapter for Visiscope SZT350
C-mount Adapter for Visiscope SZTL350
VisiCam digital cameras - 7
This is because of the lighting frequency. Extend
exposure time in the software or turn off the filament
lamp or fluorescent lamp in the environment to solve
this issue.
Qty
Cat. No.
1
630-1688
1
630-1749
1
630-1793
1
630-1793