Idiomas disponibles

E N
5 Starting the DMA
Fig. 5 - 2
14
7. Attach an adapter Luer cone (from the accessory box) to one end of the
silicone hose (see Fig. 5 - 2) and plug it in the injection adapter. Connect the
free tube end with the supplied waste vessel (see Fig. 5 - 3).
Hint:
Check the resistance of the silicone hose to the samples and cleaning agents
used.
Attaching
the
adapter
Luer cone
8. Cut a piece of approx. 250 mm length from the silicone hose contained in the
accessory box.
9. Attach the silicone hose to the air pump outlet (see Fig. 5 - 3).
10. If a printer will be used, plug the interface cable into the COM 2 connector at
the rear of the DMA.
11. Check the operating voltage.
Important:
Before switching the DMA on, make sure that the correct line voltage is
available (AC 85 to 260 V, 48 to 62 Hz). If large voltage fluctuations are to
be expected, the use of a constant voltage source (UPS) is
recommended.
The non-fused earth conductor of the power cord (or power inlet) has to
be connected to earth.
12. Connect the power inlet of the DMA to the mains using the power cord.
13. Turn on the DMA using the "POWER" switch at the rear side of the
instrument. The green light on the front indicates that the power is on. After
the start-up procedure the cell light is on continuously.
Fig. 5 - 3
Placing of the waste vessel
XDLIB28