Checking Procedure, Adjustment And Calibration; Definitions; Checking Procedure: Density Check - Anton Paar DMA 4100 Manual De Instrucciones

Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

E N

7 Checking Procedure, Adjustment and Calibration

18
"reset sample identif.": Select if the entered sample identification should be
deleted after the measurement.
"API input": Select the kind of density for calculating the API functions:
density, density (not viscosity correct.), special adjustment 0-4.
DMA 5000:
"delay": This function is only available when a temperature scan is activated.
When measurements are carried out with the temperature scan, a "delay
time" (from 0 to 9999 minutes) can be entered: After reaching the new
measuring temperature, the DMA 5000 waits for the set time before the
measurement is carried out.
"temp. scan": Measurements by different temperatures are performed
automatically. Start, end and step values can be entered.
7 Checking Procedure, Adjustment and
Calibration
7.1

Definitions

Adjustment of the density meter:
The process of bringing the instrument into a state suitable for use, by setting
or adjusting the density instrument constants.
Calibration of the density meter:
Various processes for establishing the relationship between the reference
density of measurement standards and the corresponding density reading of
the instrument.
Calibrations are performed to determine the deviation of the displayed
density values from the reference values of density standards.
7.2

Checking Procedure: Density Check

The "density check" function allows you either to check the validity of the factory
adjustment after transport or the validity of your own adjustments.
To check the factory adjustment, pure water is used as calibration fluid.
To check your own adjustments either degassed, bi-distilled water or
different density calibration fluids or standardized samples can be used.