Mask for children from 1 to 12 months of age
Ergonomic mask for smaller children:
· Its size is suitable for the facial physiology of younger children in order to limit the
dispersion of medicine on the skin compared to the use of larger paediatric
masks.
· The adjustable fitting, which connects the mask to the nebuliser cup,
enables you to keep the baby reclined. The nebuliser cup remains in a
correct position, thus obtaining maximum efficiency of the therapy.
· The mask is made of soft silicon to ensure comfort during the
therapy.
CLEANING, SANITISING AND DISINFECTING
Switch off the device before any cleaning procedure and unplug the power cable from the
socket.
DEVICE AND TUBING EXTERIOR
CLEANING
Use only a damp cloth with antibacterial soap (non-abrasive and with no solvents of any
sort).
NEBULISER CUP AND ACCESSORIES
We advise personal use of the nebuliser cup and the accessories to prevent potential risks
of transmitting infection.
Open the nebuliser cup by turning the upper part (C6.1) anticlockwise, remove the nozzle
(C6.2) by pressing as shown by the 2 arrows in the "Connection diagram" sec. C6.
SANITISATION
Before and after each use, sanitise the nebuliser cup and the accessories, choosing one of
the methods described below.
(method A): Accessories (C2-C6.1-C6.2-C6.3-C6.4-C7-C8) can be sanitised with method A
Sanitise the accessories under potable hot water (approximately 40°C) with a gentle dish
detergent (non-abrasive) or in the dishwasher with a hot cycle.
(method B): Accessories (C2-C6.1-C6.2-C6.3-C6.4-C7-C8) can be sanitised with method B
Sanitise the accessories by immersing them in a solution of 50% water and 50% white
vinegar. Lastly, rinse abundantly with hot potable water (approximately 40°C).
(method C): Accessories (C2-C6.1-C6.2-C6.3-C6.4-C7-C8) can be sanitised with method C
Sanitise the accessories by boiling in water for 20 minutes. It is preferable to use de-
mineralised or distilled water to prevent calcium deposits.
After having sanitised the accessories, shake them vigorously and place them on a
paper towel. Alternatively, dry them with a jet of hot air (for example, a hair dryer).
DISINFECTION
Accessories (C2-C6.1-C6.2-C6.3- C6.4-C7-C8) can be disinfected
The disinfection procedure described in this paragraph is to be followed before using
the accessories. It is effective on the parts that undergo this treatment only if each
step is followed and only if the parts to be treated have previously been sanitised. The
disinfectant must be an electrolytic chloroxidizer (active principle: sodium hypochlorite)
specific for disinfecting, which is available in any pharmacy.
Implementation:
- Fill a container big enough to hold all of the parts to disinfect with a solution of potable
water and disinfectant, according to the proportions indicated on the packaging of the
disinfectant.
- Completely immerse each part in the solution, taking care to avoid the formation of air
bubbles on the parts. Leave the parts immersed for the amount of time indicated on the
packaging of the disinfectant associated with the concentration chosen for the solution.
- Remove the disinfected parts and rinse abundantly with warm potable water.