Safety Considerations

Read and understand this manual completely before attempting to set up or
use this instrument.
This equipment has been designed and tested to conform to IEC1010-1 safety standards, as
applicable to laboratory instrumentation. This applies only to the FB200 when used as specified
in the documentation, in its intended applications, with Fisher Biotech approved electrophoresis
apparatus only. Usage in any other manor may not provide similar performance or safety
protection.
This equipment is provided with a 3-conductor, grounded AC line cord. The protective earth
ground is necessary for safe operation. Do not use any other AC line cord with this instrument.
The FB200 is a high voltage power supply capable of generating dangerous levels of voltage and
current during operation. Exercise caution when working around and with the electrical
connections of this equipment. Always check electrical connectors, wires, and associated
apparatus for any signs of wear or damage before using with this equipment. Be sure to use only
electrophoresis equipment that is suitably rated for the voltage and current capabilities of the
FB200 power supply.
The output of the FB200 power supply is intended for connection to electrically isolated
electrophoresis apparatus only. Use only with electrically isolated electrophoresis apparatus with
minimum isolation of 600V. Do not connect any terminal of the FB200 output to earth ground.
This may impair the safety protection provided by the equipment, or cause equipment damage.
The high voltage output of the FB200 power supply takes some amount of time to decay when
unloaded or lightly loaded. Wait a minimum of 60 seconds after stopping a run before touching
the power supply leads.
This equipment has a protective ground leakage current of 0.5mA at 120VAC, and 1mA at
230VAC. While significantly below the 3.5mA limitation for laboratory instrumentation, some
applications and locations require a leakage current below 0.5mA. Check the specific
requirements of your application before using this equipment.
This equipment is for indoor use only.
FB200-2 MANUAL
Page 2
7608175 Rev A
8/02