Battery Replacement - GPI 09 Computer Electronics Guía De Operaciónes

Ocultar thumbs Ver también para 09 Computer Electronics:
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

Battery Replacement

The computer display is powered
by two 3-volt lithium batteries which
may be replaced while the meter
is installed. Battery life is 5 years.
When batteries are removed or lose
power, the Batch and Cumulative
Totals and the field and factory
calibrations are retained. If the
display becomes dim, blank or the
low battery message appears (see
below), replace batteries as follows:
To replace batteries, remove the
corner screws from the face of
the meter and lift off the computer
assembly. Remove the old batteries
and clean any corrosion from the
terminals. Install new batteries.
Make sure the positive post is in
the correct position. When the
batteries are replaced, the faceplate
will power ON. Check the display
to ensure normal functions have
resumed before assembling again.
Make sure the seal is fully seated.
Put the computer assembly back
on the turbine and secure with the
corner screws.
TROUBLESHOOTING
A. METER IS NOT ACCURATE.
1. Field Calibration not per-
formed properly. Field cali-
brate again or select Factory
Calibration. See Operation
and Calibration instructions.
2. Factory Calibration not
suitable for liquid being
measured. Perform a Field
Calibration according to
Calibration instructions.
3. Meter operated below mini-
mum flowrate. Increase
flowrate.
4. Meter partially clogged with
dried liquid. Remove meter.
Clean carefully.
5. Meter bearings partially
clogged with dried liquid.
R e m o v e m e t e r. C l e a n
carefully.
6. Sealant material wrapped
around rotor. Remove meter.
Clear material from rotor.
Make sure rotor spins freely.
7. I n s t a l l e d t o o c l o s e t o
fittings. Install correctly. See
Installation instructions.
8. Installed too close to motors
o r el e ct ri c al l y " n o i s y"
environment. Install correctly.
See Installation instructions.
B. READOUT FADED OR BLANK.
1. Batteries weak, dead, or
not connected. Remove
c o m p u t e r a n d r e p l a c e
batteries. See Maintenance
section.
2. Computer defective. Contact
local dealer or distributor.
C. NORMAL FLOWRATE BUT
METER DOES NOT COUNT.
1. F i e l d C a l i b r a t i o n n o t
performed correctly. Field
Calibrate again or select
Factory Calibration. See
Operation or Calibration
sections.
2. Rotor stuck or damaged.
Remove meter. Make sure
rotor spins freely.
3. Sealant material wrapped
around rotor. Remove meter.
Clear material from rotor.
Make sure rotor spins freely.
4. Computer defective. Contact
your local dealer or distributor.
D. REDUCED FLOWRATE AND
METER DOES NOT COUNT.
1. Meter clogged with dried
liquids. Remove meter. Clean
carefully. See Maintenance
section. Make sure rotor spins
freely.
5