TROUBLE SHOUTING

! IMPORTANT
Problem
No power
Digital display has
no image
Show start up of
device
Abnormal noise or
vibration
during
operation

TECHNICAL DATA

Height x wide x deepth:
Weight:
Tension:
Power:
Timer:
Memory:
Speed:
16
If you find a problem different from the following, contact your distributor
and send the unit to repair.
Cause
Bad connection of power.
Fuse burn out.
Defective PC board.
Defective driving motor or gear.
Cover not fully closed.
Defective driving motor or gear.
Defective starting condenser.
Too much alginate and/or water.
Mixing bowl is empty.
Driving gear damage.
Alginate powder spillage because of
loose cap.
305 x 205 x 245 mm
18 kg
AC 230 V, 50/60 Hz
400 W
1-16 seconds
3 memories
3600 r.p.m.
Solution
Inspect the connection of power.
Inspect fuse and change if needed.
Return to dealer for service.
Return to dealer for service.
Make sure cover of mixer is securely
closed.
Return to dealer for service.
Return to dealer for service.
Maximum amount of alginate is 100
grams (including water + alginate).
Never use unit with an empty mixing
bowl.
Return to dealer for service.
Secure bowl cap and tighten.