Guidance and declaration of manufacturer-electromagnetic emissions
The device is intended for use in the electromagnetic environment specified below. The customer or the user of the device should assure that it is
used in such an environment.
Emissions test
RF emissions
CISPR 11
RF emissions CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations/ flicker emissions
IEC 61000-3-3
The device is intended for use in the electromagnetic environment specified below. The customer or the user of the device should assure that it is
used in such an environment.
Immunity test
Electrostatic
discharge (ESD)
IEC 61000-4-2
Electrostatic
transient/burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Voltage dips,
short interruptions
and voltage variations
on power supply input
lines
IEC 61000-4-11
Power frequency (50/60 Hz)
magnetic field
IEC 61000-4-8
Compliance
Group 1
Class B
N/A
N/A
Immunity aspects
IEC 60601 test level
± 6 kV contact
± 8 kV air
± 2 kV for power supply
lines
± 1 kV for input/output
lines
±1kV differentail mode
±2kV common mode
< 5% UT
(>95% dip in UT)
for 0.5 cycle
40% UT
(60% dip in UT)
for 5 cycle
70% UT
(30% dip in UT)
for 25 cycle
<5% UT
(>95% dip in UT)
for 5 secretary
3 A/m
Electromagnetic environment – guidance
The device uses RF energy only for its internal function. Therefore,
its emissions are very low and are not likely to cause any
interference in nearby electronic equipment.
The device is suitable for use in all establishments, including
domestic establishments and those directly connected to the public
low-voltage power supply network that supplies buildings used for
domestic purposes.
Electromagnetic environment –
Compliance level
guidance
Floors should be wood, concrete or ceramic
± 6 kV contact
tile. If floors are covered with
± 8 kV air
synthetic material, the relative humidity
should be at least 30 %.
N/A
N/A
N/A
Power frequency magnetic fields should be
3 A/m
at levels charactertic of a typical location in a
typical commercial or hospital environment.