CORRECTIVE AND PREVENTIVE MAINTENANCE AND
PRESERVATION

Disinfection

- Use clean and soft cloth dampened in alcohol 70% to disinfection of the equipment.
- Never use corrosive disinfectants or solvents.
Note: Use gloves and other
systems of protection,
during the disinfection.

Preventive Maintenance

The equipment should be calibrated routinely, as per the legislation in force in the
country.
But never with a period exceeding 3 years. In order to protect your equipment, seek
Gnatus technical assistance for periodic revisions of preventive maintenance.

Corrective Maintenance

If the equipment has any abnormality, check if the problem is related to any of the items
listed in the item Unforeseen Events (situation, cause and solution). If it is not possible to
solve the problem, turn off the equipment, and request Gnatus technical assistance.

Replacement of fuses

Remove the power input cable (04) (D).
With the help of a screwdriver, release the fuseholder 's cover (E), after wards replace
the fuse.
WARNING: After replacing the fuses for the first time, we recommend to reserve some
fuses for future replacements.