How to Test Glucose Control
Use ONLY TRUEcontrol Glucose Control Solution
with the TRUEtrack Meter
1. Wash hands with soap and water, dry
thoroughly. Allow Control, vial of Test Strips
and Meter to adjust to room temperature
(68-77°F). Write date first opened on Control
Label and/or Test Strip vial label.
2. Check dates on Control label and Test Strip
vial label. Do not use Control if 3 months past
written opened date or past date printed next
to EXP, whichever comes first. Do not use Test
Strips 120 days past written opened date or
past date printed next to EXP, whichever comes
first. Discard out-of-date products and use new
products if either date has passed.
3. Swirl or invert bottle gently to mix
Control. DO NOT SHAKE!
4. Remove Test Strip from vial.
Recap vial immediately.
Note: Use Test Strip quickly after removal
from vial. Test Strips that have been
left out of vial too long will give
an error message. If error message
appears in Display, discard old Test
Strip and test with new Test Strip.
5. Insert Test Strip Contact End
(Contacts facing up) into Test Port.
Meter turns on. Keep Test Strip in
Meter until testing is finished.