WARNING: Upon opening the test strip carton,
examine the product for missing, damaged or broken
parts. Ensure the test strip vial cap is securely closed.
If the product is damaged or the vial cap is not closed,
DO NOT use the test strips for testing; product
may give inaccurate results. Contact Trividia Health
Customer Care at 1-800-803-6025 for replacement
and assistance.
Health Related Information
For Healthcare Professionals:
If comparing results between TRUEtrack and a lab
system, perform a TRUEtrack blood test within 30
minutes of lab test. Results from TRUEtrack are
considered accurate if within +/– 20% of lab results.
5
If patient has recently eaten, fingerstick results from
TRUEtrack can be up to 70 mg/dL higher than
venous lab results.
Do not use venous blood with
6
TRUEtrack.
Instructions for use of the System for more
than one person may be found in the Quality
Assurance/Quality Control Manual. The QA/QC
Manual is available, call 1-800-803-6025 or
1-954-677-4599.
6