Overload Protection:
600V RMS
• Voltage:
2000µA (leads), 1000A (clamp)
• Current:
AC Current Measurement
Range
600µA
2000µA
60A
600A ~ 1000A
Overload Protection:
600V RMS
• Voltage:
2000µA (leads), 1000A (clamp)
• Current:
Frequency:
45 to 400Hz
Minimum Current for Frequency Measurement:
Response:
True RMS
Resistance Measurement
Range
Resolution
600Ω ˜ 6MΩ
0.1Ω ˜ 0.001MΩ
60MΩ
0.01MΩ
Overload Protection:
Capacitance Measurement
Range
60nF ˜ 6000µF
Overload Protection:
Frequency Measurement
Range
99.99Hz ˜ 499.9kHz
Overload Protection:
Duty Cycle Measurement
Range
Resolution
0.1 ˜ 99.9%
0.1%
Overload Protection:
Diode Test
Overload Protection
600V RMS
Continuity Test
Overload Protection
600V RMS
Non Contact Voltage Detector
On Voltage
Appx. 25V AC and above
Resolution
0.1µA
1µA
0.01A
0.1A ~ 1A
Accuracy
± (1.0% + 4 digits)
± (2.0% + 4 digits)
600V RMS
Resolution
0.01nF ˜ 1µF
600V RMS
Resolution
0.01Hz ˜ 100Hz
600V RMS
Accuracy
± (0.2% per kHz + 0.1% + 5 digits)
600V RMS
Range
Test Current
6.0V
Appx. 0.25mA
Open Circuit Voltage
< 0.44V
Accuracy
± (2.0% + 5 digits)
± (1.5% + 5 digits)
± (2.9% + 15 digits)
± (1.9% + 8 digits)
400µA (leads) or 20A (clamp)
Accuracy
± (3.5% + 6 digits)
Accuracy
± (0.1% + 4 digits)
Open Circuit Voltage
< 1.8V DC
Threshold (Appx.)
~40Ω