SYMPTOM
Blank display when the
balance is turned on.
Unstable display
The balance doesn´t turn on
and turn off
The balance doesn´t tare
The recharging indicator
doesn´t light on.
Reading display unstable.
When the load is removed,
the display doesn´t remain at
zero.
Serious error
Battery can not be recharged Defective battery
Short remote control
distance
CAUSE
Defective battery
Low battery
Defective
key
Defective
key
Defective AC/Adaptor
Unstable load.
The balance has been for a
long time in a wet
environment.
Insuficient warm up time.
The crane has not been
hung properly.
Defective Outlet
The remote sensor is dirty
SOLUTION
Change the battery
Discharged battery
Recharge the battery
Check and clean it.
Check and clean it.
Check the AC/Adaptor
The AC/Adaptor is not
connected properly.
Stabilize the load
Defective loadcell cable.
Place the balance in a dry
environment.
Give the scale longer
warm-up time, around 3-5
minutes
The crane scale has been on
the ground for a long period
of time.
Check the balance and hook.
Change the battery
Change the Outlet
Check and clean it.
Remote control batteries are
dead
10