Cancel Button - Lantheus Medical Imaging VIALMIX Guia Del Usuario

Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

C A N C E L B U T T O N
Vialmix
is equipped with a button that allows the user to terminate the
®
activation cycle. By pressing the CANCEL button after the shaking cycle
is initiated, the motor shuts down and the shaking cycle stops.
The CANCEL button is available in the unlikely event that the activation
cycle must be stopped for safety reasons (for example, should the vial
become dislodged or a mechanical failure of Vialmix
The following table shows the LED message when the CANCEL button is
pressed, the associated Vialmix
and the action required of the user:
Associated
LED
Message
"EE" alternating
Stops shaking,
with "CL"
emits an audible
alarm
response, the likely cause of the message
®
Vialmix
Likely Cause
®
Response
of Message
CANCEL
button was
pressed during
operation.
occur).
®
Action
Required by
User
Power off the
u n i t . Do not
use the Definity
®
vial since it was
not acceptably
a c t i v a t e d . R e t a i n
the vial and
call Te c h n i c a l
S e rv i c e s .
(See pg. 1 0 . 1 )
S A F E T Y F E AT U R E S
■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■
IMPORTA N T:
Pressing the

CANCEL button

will terminate the
activation cycle
and render Definity
®
unacceptable for
use. Retain the
vial and call
Technical Services
for assistance.
(See pg 10.1)
4.2