T R O U B L E S H O O T I N G
■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■
■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■
IMPORTA N T:
If you need
assistance, or have
technical questions
about Vialmix
®
, call
Technical Services.
(See pg 10.1)
6.2
5 1 5 3 7 0 - 0 8 1 0
LED
Message
Likely Cause
LED display does
No main power
not operate
when power
switch is on.
Fuse is blown.
Motor does not
Cover is open.
operate after
button
START
Motor is
is pressed.
overheated.
Cover has been
Vialmix
stops
®
opened.
shaking during a
cycle.
Unit deviated
from acceptable
shaking rate.
Fuse is blown.
Motor is
overheated.
Corrective Action Required
Check power supply and powe r
c o rd connection.
Contact Technical Services for
a s s i s t a n c e,or re fer to page 8.1 fo r
instructions to change the fuse.
Close the cove r.
A l l ow the motor to cool dow n .
Do not attempt to use Definity
vial since it was not acceptably
a c t i v a t e d . Retain the vial and call
Technical Serv i c e s . (See pg 10.1)
Do not open the cover during
a c t i v a t i o n .
Do not use the Definity
®
it was not pro p e r ly activated.
Retain the vial and call Te c h n i c a l
S e rv i c e s . (See pg 10.1)
Do not use the Definity
vial since
®
it was not pro p e r ly activated. R e t a i n
vial and call Technical Services fo r
a s s i s t a n c e.(See pg 10.1) See page
8.1 for instructions to change fuse.
A l l ow motor to cool dow n . D o
not use the Definity
vial since it
®
was not pro p e r ly activated. R e t a i n
vial and call Technical Services fo r
a s s i s t a n c e.(See pg 10.1)
®
vial since