Sophysa PRESSIO 2 PSO-4000 Manual De Instrucciones página 16

Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 65
Display
Description
Rapid scrolling of the ICP curve from right to left.
Hiding of the ICP curve from the screen.
Slow scrolling of the ICP curve from right to left.
In this mode, the speeds offered by the
suitable for slow scrolling.
The display of a new ICP curve takes place 10 minutes
after the start of monitoring. During this time the button
is greyed out and not functional.
5.4.3.2. Scrolling Speed for the ICP curve
When the ICP curve is displayed on the screen, the
button enables changes to the scrolling speed of the ICP
curve.
To change scrolling speed, press the
speed from the other settings available.
NOTE
When the MEAN scrolling mode is activated, other val-
ues are displayed suggesting slower scrolling.
The scrolling speeds are expressed in millimeters per second
(mm/s) or in centimeters per hour (cm/h).
5.4.3.3. Pressure Scale for the ICP curve
When the ICP curve is displayed on the screen, the
button enables changes to the scale for the height of the
curve.
To change the pressure scale, press the
select a scale from the settings available.
The pressure scale intervals are expressed in millimeters of
mercury (mmHg).
Select the AUTO option to obtain the pressure scale setting
most suitable for the values measured among the settings
available.
5.4.3.4. Event Entry
Press the
button to display a virtual keyboard, and to
enter an event.
Each event has a maximum length of 50 characters. It is dat-
ed and added to the history.
The entered events cannot be changed or erased.
To add more information to an existing event:
1.
Enter the history mode.
2.
Place the cursor close to the event already created.
3.
Add a new event.
Page 16 of 174
5.4.3.5. Access to the History
Press the
ing in progress.
5.4.3.6. Access to other configuration settings
Press the
Display language
Press "Language" to change the language used on the
button are
screen.
Alarms
Press "Alarm" to access alarms activation/deactivation and
alarm thresholds limits for ICP and ICT.
Recording modes
Press "Recording mode" to manage the history recording
modes.
Degrees Celsius or Fahrenheit
Press "Parameters", then "Temperature unit" to change the
units used for displaying temperature (C° or F°).
button and select
Time zone
Press "Parameters" then "Time Zone", to select the time
zone where the Monitor is to be used.
Screen brightness
Press "Parameters", then "Brightness" to adjust the screen
brightness.
5.4.3.7. Locking and Unlocking the Action Bar
Press and hold the
to access the configuration settings menu.
The action bar re-locks:
– by pressing the
button, and
– automatically after 5 minutes of inactivity.
5.4.3.8. Temporary Audible Alarm Suspension
The
triggered.
Press this button to suspend audible alarms temporarily.
Audible alarms reactivate automatically after 2 minutes.
When audible alarms are suspended, the
placed by the following button in the bottom right corner of
the touch screen:
Audible alarms can be reinitialized by changing the value of
one of the ICP or ICT alarm thresholds, or by deactivating
and reactivating the physiological alarms.
5.4.4. Alarms
The physiological alarms trigger when the ICP and/or ICT
values exceed the alarm thresholds set.
button to consult the history of the monitor-
button to access the following settings:
button to unlock the action bar and
button,
button appears automatically when an alarm is
button is re-