E1
Upper sensor error
E2
Upper alarm sensor error
E3
Lower sensor error
E4
Lower alarm sensor error
E5
Ambient sensor error
E6
Evaporator sensor error
E7
Upper condenser sensor error
E8
Lower condenser sensor error
Do not put the items in product that have just been connected to the power supply. Allow the
empty product to run for a period of time (approximately 24 hours) before storing the items in the
cabinet.
When placing items, if the items are too dry or too wet, it will affect the change of humidity inside
the cabinet, and the items should be sealed and stored. The humidity of the working environment of
the product will also affect the humidity in the cabinet, especially when the door is opened too much
or not closed. Each time the item is stored, it cannot exceed 1/3 of the volume of the cabinet. After
reach the actual temperature and operate normally for 24 hours, it can be placed in another 1/3 of
the volume of the cabinet.
Be careful not to block air outlet and air inlet when store items. Do not put the items directly on
the bottom of the refrigerating chamber, and should be placed on the bottom shelf, otherwise it will
affect the refrigeration effect of the refrigerating chamber. Do not open the door during the cooling
process, otherwise it will cause the temperature to rise. Due to the cooling inertia, there may be a
certain difference between the display temperature and the set temperature when the product
reaches the set value. This is a normal phenomenon. Note: Children are not allowed to play with
refrigerators.
7. Defrost, Stop Using and Maintenance
Warning:
To prevent electric shock or injury to persons, be sure to cut off the power before performing any
repairs and maintenance on the product. Do not inhale the medicine or suspended particles around
1.
Check the upper sensor connection
2. Replace the upper sensor
1.
Check the upper alarm sensor connection
2. Replace the upper alarm sensor
1.
Check the lower sensor connection
2. Replace the lower sensor
1.
Check the lower alarm sensor connection
2. Replace the lower alarm sensor
1.
Check the ambient sensor connection
2. Replace the ambient sensor
1.
Check the evaporator sensor connection
2. Replace the evaporator sensor
1.
Check the upper condenser sensor connection
2. Replace the upper condenser sensor
1.
Check the lower condenser sensor connection
2. Replace the lower condenser sensor