5HFRPPHQGDWLRQV IRU XVH
-
Use the level of currently strictly necessary in relation to the
operational conditions and the conditions of the patient's tissue,
keeping the knife in constant movement without exerting pressure.
-
Keep the knife electrode tip clean.
-
The CUT setting delivers higher current to allow cleaner incisions; the
COAG setting is suitable for working on haemorrhagic areas or
bleeding tissue in general.
-
If you find that cutting power is insufficient or if fragments of tissue
tend to adhere to the knife electrode, increase output power by turning
potentiometer (5) clockwise.
-
Sparking in the area of the electrode indicates that output power is too
high; in this case reduce power by turning potentiometer (5) counter-
clockwise.
:DUQLQJV
BEFORE USING THE APPLIANCE CHECK ITS COMPATIBILITY WITH
THE OTHER EQUIPMENT/APPLIANCES IN USE IN THE AREA; THE
ELECTROSURGICAL
DISTURBANCES DURING NORMAL OPERATION WHICH COULD
INTERFERE WITH THE CORRECT OPERATION OF OTHER DEVICES
IN THE AREA.
ALWAYS SWITCH OFF THE APPLIANCE WHEN YOU FINISH
WORKING
WITH
IRRESPECTIVE OF THE MOTIVE.
IF
YOU
NOTICE
IMMEDIATELY AND DISCONNECT IT FROM THE ELECTRICAL
POWER SUPPLY. BEFORE RECONNECTING THE POWER SUPPLY
FIND THE CAUSE OF THE FAULT AND REMEDY IT. IN THE CASE OF
PROBLEMS, CONSULT CARLO DE GIORGI.