Owner's Guide
B. No water at insert
1. Assure that the H
O control is properly adjusted. A fine
2
mist or rapid drip of water should be observed at the tip of
the insert.
2. Check that the main shut-off valve is open.
3. Check the disposable in-line water filter for possible
clogging.
4. Check the recommended customer supplied in-line water
filter for possible clogging.
5. Make sure that the water lines are not kinked.
6. Make sure that the water knob is turned to the "open"
position (clockwise).
7. Check the insert tip for possible clogging.
C. Unit sprays water, but the same insert does not vibrate
• Try a different insert. Sometimes a worn out and/or damaged
insert is not detectable to the naked eye.
• Check indicator light to see if unit is set to the same frequency
as the insert. If not, move switch on the back of the unit to the
correct frequency.
X. Specifications
A. Unit
Catalog No.
Voltage
US100R115M
115V
US100R115T
115V
US100R230CE
230V
US100R230UK
230V
US100R240
240V
Operating Frequency:
25/30 kHz
Power Consumption:
40–70 W
Overall Dimensions:
5,5 x 9 x 6 inches
(13.97 x 22.86 x 15.24 cm)
Weight:
10lbs (4.54kg)
Fuse Type:
5 x 20 mm, Slo-Blo Fuses
115V units
230–240V units 0.5 Amp, 250 V T-Type
B. Insert
Catalog No.
Frequency
US1025K
25 kHz
US1030K
30 kHz
US1025KSP
25 kHz
US1030KSP
30 kHz
US100025K
25 kHz
US100030K
30 kHz
US1025KOM
25 kHz
US1030KOM
30 kHz
Frequency
Power
50/60 Hz
100W max
50/60 Hz
100W max
50/60 Hz
100W max
50/60 Hz
100W max
50/60 Hz
100W max
1.0 Amp, 250 V T-Type
Catalog No.
Frequency
US1025KOS
25 kHz
US1030KOS
30 kHz
USG1025K
25 kHz
USG1030K
30 kHz
USG1025KSP
25 kHz
USG1030KSP
30 kHz
USG1025KOS
25 kHz
USG1030KOS
30 kHz
C. Explanation of Symbols
Protective Earth Ground
Caution/Warning
Type BF Applied part
Footswitch
Water Connection
Fuse
Output Intensity Adjustment
Refer to Instruction Manual
Burn Hazard
Electric Shock Hazard
Explosion Hazard
Temperature Limitation
Fire hazard
%
Humidity Limitation
Atmospheric Pressure Limitation
7