Maintenance And Cleaning; Routine Maintenance; Cleaning And Disinfection; Service Life - GCE Zen-O lite RS-00600 Manual Del Usario

Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 176

11 . MAINTENANCE AND CLEANING

11 . 1 . ROUTINE MAINTENANCE

WARNING: Do not use lubricants on this device or any of its accessories .
CAUTION: Replace the cannula on a regular basis. Check with your distributor or clinician
to determine how often the cannula should be replaced.
Device will indicate with an alarm when service is required. (Also, see Chapter 10 - Trou-
bleshooting.)
EN
No special maintenance needs to be carried out by the user. Contact your provider or
manufacturer for help, in setting up, maintenance, or to report unexpected errors.

11 .2 . CLEANING AND DISINFECTION

WARNING: Do not submerge this device in liquid . Do not expose to water or precipita-
tion . Do not expose to dusty conditions .
CAUTION: Do not use cleaning agents other than those specified in this manual. Allow
the cleaning solution to dry from the cleaned surface before use.
CAUTION: Always disconnect power and turn off this device before cleaning.
Clean the exterior with a soft cloth slightly dampened with soapy water or with anti-bac-
terial wipes (Isopropyl alcohol 70% solution).
For disinfecting, use a MadaCide-FDW-Plus wipe or equivalent and follow the manufac-
turer's instructions. (Manufacturer - Mada Medical Products Inc., www.madamedical.com)
Important: The device should receive an external cleaning weekly, accessories should
be cleaned as needed. The device exterior and accessories should be cleaned and dis-
infected and the patient filter replaced prior to delivering to a new patient. The device
may become hot after operation, take additional care when replacing filters.
Nasal cannula: Refer to the original manufacturer's instruc-
tions for cleaning the nasal cannula.

11 .3 . SERVICE LIFE

The expected service life of the device is 5 years, except for
the sieve modules. The service life of the sieve modules will
depend on the operating conditions. Replace them as need-
ed, indicated by the check vents alarm. If intake and exhaust
vents are not blocked and the check vents alarm persists,
contact your distributor for instructions on replacing the sieve
modules.
Fig. 8
29/340

Publicidad

Tabla de contenido

Solución de problemas

loading

Productos relacionados para GCE Zen-O lite RS-00600

Este manual también es adecuado para:

Zen-o lite rs 00600c

Tabla de contenido