MANUFACTURER'S GUIDE AND DECLARATION - ELECTROMAGNETIC IMMUNITY
The clinical thermometer JC-132G (FDTH-V0-7) can be used in the following electromagnetic
environment. The customer or user of the clinical thermometer JC-132G (FDTH-V0-7) must
make sure that the device is used in said environment.
EMISSION TEST
Electrostatic discharges (ESD)
IEC/EN61000-4-2
Magnetic field
IEC/EN 61000-4-8
Irradiated immunity
IEC/EN61000-4-6
COMPLIANCE
± 6kV in contact
± 8kV in the air
3 A/m
3 V/m 80 MHz at 2.5 GHz
(in appliances that are not
life equipment)
12
GUIDE TO THE ELECTROMAGNETIC
ENVIRONMENT
The floors should be made of
wood, cement or baked clay. If
the floors are covered in synthe-
tic material, relative humidity
should not exceed 30%.
The magnetic field should
be the typical magnetic field
of a commercial or hospital
environment.