LEM PR1200 ACI Manual De Uso página 9

Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

IP rating according to IEC 529
IP40 with jaws closed
IP30 with jaws opened
Safety Standards
IEC61010-1 : 1993. General Requirements.
Safety requirements for electrical equipment for meas-
urement, control and laboratory use.
IEC61010-2-032: 1994. Particular requirements for
hand held current clamps for electrical measurement
and test.
IEC61010-2-031: 1993. Particular requirements for
hand held probe assemblies for electrical measure-
ment and test.
For input and output, 600 V RMS or DC, over voltage
Cat. III, Pollution degree 2
EMC Standards
RF Susceptibility
EN 50082-1: 1992 3 V/m Residential, Commercial
and Light Industry
RF Emissions
EN 50081-1: 1992 Residential, Commercial and
Light Industry
FCC Part 15 Class B
This product is designed to be safe under the follow-
ing conditions :
- indoor use
- altitude up to 2000 m
- temperature -10°C to +55°C
- maximum relative humidity 85% for tempera-
tures up to 31°C decreasing linearly to 40% relative
humidity at 50°C
Use of the probe on uninsulated conductors is lim-
ited to 600 V RMS or DC and frequencies below 1
kHz.
Safety in its use is the responsibility of the operator
who must be a suitably qualified or authorised person.
Do not use the probe if any part of the probe including
the lead or connector(s) appear to be damaged or if a
malfunction of the instrument is suspected.
When using the probe ensure that your fingers are
7

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

Pr1200 acvPr1201 aciPr1201 acvPr1200 x

Tabla de contenido