Wiring Diagram For Operating Mode 2 - Sartorius YSB01 Guia De Instalacion

Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

Wiring Diagram for Operating Mode 2

To operate the relay box in Operating Mode 2 (encoded data), DIP switches 1 through
3 and 5 through 7 must be open ("Off") and DIP switch 4 closed ("On").
Equal to (=) + Greater than (>)
B6
B5
B4
B3
B2
B1
Combics application: "Classification" (5 classes)
Class 1:
Class 2:
Class 3:
Class 4:
Class 5
Ready/Set
8
A10
Ground (customer installation)
A9
Supply voltage (customer installation)
A1
UNI-IN
Less than (<)
Less than (<) + Equal to (=)
Equal to (=)
Greater than (>)
Set
8
Ground (from control device)
UNI
MINOR: Less than (<)
PARES: Equal to (=)
MAJOR: Greater than (>)
SET
Control signal: Less than (<)
Control signal: Less than (<) + Equal to (=)
Control signal: Equal to (=)
Control signal: Equal to (=) + Greater than (>)
Control signal: Greater than (>)
Control signal: Set
2
3
4
5
6
Power supply
(from control device)
Ground
(from control device)
C1
C3
Relay 2
Relay 3
Relay 4
Relay 5
Relay 6
Relay 8