3. Insert the probe VERY gently and slowing into the ear canal.
Hold the thermometer so the probe faces straight
in the direction of the eardrum, Fig 3.
It is important to point the probe tip toward the
eardrum and is inserted inside the ear canal until
slight resistance is felt.
Hold the outer ear and gently pull it up and toward
the back of the head to straighten the ear canal,
Fig 4 & Fig 5.
Always insert the thermometer into the same ear
with the same direction and depth.
4. Press and hold the SCAN button. The thermometer take a
reading and make a 'beep' sound, indicating the measure-
ment is complete.
5. Remove the thermometer from the ear canal. The
thermometer will display the result on the backlit LCD
screen.
NOTE: After each measurement "
display above the measurement. Once the
"
" disappears (approx. 6 seconds), a new
measurement can be taken, Fig 6.
6. After the test is complete, press the M (Memory)
button or wait 1 minute for the display to
return to standby mode and display the
ambient temperature. After 1 minute in standby
mode, the thermometer will automatically
shut OFF.
NOTE: The LED light between the Memory and Settings button will illuminate
according to the temperature reading.
• GREEN = 93.2°F ~ 100.3°F (34.0°C ~ 37.9°C)
• RED = 100.4°F ~ 109.3°F (38.0°C ~ 42.9°C)
• LO = < 89.6°F (32.0°C)
• HI = > 109.3°F (42.9°C)