complaints, press the START/STOP button to release
the air immediately from the cuff.
Loosen the cuff and remove it from your wrist.
• If the cuff pressure reaches 40 kPa (300 mmHg),
the unit will automatically deflate. Should the cuff
not deflate when pressures reaches 40 kPa (300
mmHg), detach the cuff from the wrist and press
the START/STOP button to stop inflation.
• Before use, make sure the device functions safely
and is in proper working condition. Check the
device, do not use the device if it is damaged in any
way. The continuous use of a damaged unit may
cause injury, improper results, or serious danger.
• Do not wash the cuff in a washing machine
or dishwasher!
• The service life of the cuff may vary by the
frequency of washing, skin condition, and storage
state. The typical service life is 10000 times.
• It is recommended that the performance should
be checked every 2 years and after
maintenance and repair, by retesting at least the
requirements in limits of the error of the cuff
pressure indication and air leakage (testing at least
at 50mmHg and 200mmHg).
• Please dispose of ACCESSORIES, detachable
parts, and the ME EQUIPMENT according to the
local guidelines.
• Manufacturer will make available on request
circuit diagrams, component part lists,
descriptions, calibration instructions,etc., to assist
to service personnel in parts repair.
• The operator shall not touch output of batteries
and the patient simultaneously.
• Cleaning :Dust environment may affect the
performance of the unit. Please use the soft cloth to
clean the whole unit before and after use. Don't use
any abrasive or volatile cleaners.
• The device doesn't need to be calibrated within
two years of reliable service.
• If you have any problems with this device, such
as setting up, maintaining or using, please contact
the SERVICE PERSONNEL of HoMedics. Don't open
or repair the device by yourself in the event of
malfunctions. The device must only be serviced,
repaired and opened by individuals at authorized
sales/service centers.
• Please report to HoMedics if any unexpected
operation or events occur.
• Keep the unit out of reach of infants, young
children or pets to avoid inhalation or swallowing of
small parts. It is dangerous or even fatal.
• Be careful to strangulation due to cables and
hoses, particularly due to excessive length.
• At least 30 min required for ME equipment to
warm from the minimum storage temperature
between uses until it is ready for intended use. At
least 30 min required for ME equipment to cool from
the maximum storage temperature between uses
until it is ready for intended use.
• This equipment needs to be installed and put into
service in accordance with the information provided
in the ACCOMPANYING DOCUMENTS;
• Wireless communications equipment such as
wireless home network devices, mobile phones,
cordless telephones and their base stations, walkie-
talkies can affect this equipment and should be
kept at least a distance d away from the equipment.
The distance d is caculated by the MANUFACTURER
from the 80MHz to 5.8 GHz column of Table 4 and
Table 9 of IEC 60601-1-2:2014, as appropriate.
• Please use ACCESSORIES and detachable parts
specified/ authorised by
MANUFACTURER. Otherwise, it may cause damage to
the unit or danger to the user/patients.
• There is no luer lock connectors used in the
construction of tubing, there is no possibility
that they might be inadvertently connected to
intravascular fluid systems, allowing air to be
pumped into a blood vessel.
• Please use the device under the environment
which was provided in the user manual. Otherwise,
the performance and lifetime of the device will be
impacted and reduced.